Features
      ・   350 ohm precision calibrator 
・   Output calibrated by connecting this unit with a digital indicator, etc., for the strain gage based transducer.
Specification
| Specification name | Specification contents | 
|---|---|
| Rated output | 0.5 mV/Vrange 10 steps  2 mV/V range 10 steps 5 mV/V range 10 steps 10 mV/V range 10 steps  | 
            
| Error | 0.02 %F.S. (Temperaturerange from 15℃ to 25℃) | 
| Excitation, maximum | 12 VDC | 
| Zero balance | ±0.001 mV/V | 
| Input resistance | 350 Ω ±7 Ω | 
| Output resistance | 350 Ω ±7 Ω | 
| Temp. effect on zero | 0.03 %F.S. /10℃ | 
| Temp. effect on output | 0.03 %F.S. /10℃ | 
| Operating temp./humidity range | Temperature 0℃ to50℃,  Humidity 85 %RH or less(Non condensing)  | 
            
| Outline dimensions(WxHxD) | 185mm x 47mm x 125mm (excludes protruding parts) | 
| Weight | Approx.1.3 kg | 